NO | 76 |
---|---|
Acronym | UYRK |
Name | University of York |
Address 1 | |
Address 2 | NCAS, University of York |
Address 3 | York YO10 5DD United Kingdom |
Country/Territory | United Kingdom of Great Britain and Northern Ireland |
Website | http://www.york.ac.uk/ |
Name | Katie Read |
---|---|
Prefix | Dr. |
katie.read@york.ac.uk | |
Organization No | 76 |
Organization acronym | UYRK |
Organization name | University of York |
Organization country/territory | United Kingdom of Great Britain and Northern Ireland |
Address 1 | NCAS, Department of Chemistry |
Address 2 | University of York, Heslington |
Address 3 | York YO10 5DD United Kingdom |
Country/territory | United Kingdom of Great Britain and Northern Ireland |
Tel | +44-1904-322565 |
Fax | +44-1904-322516 |
Last updated date | 2024-04-04 |
Name | Lucy Carpenter |
---|---|
Prefix | Prof. |
lucy.carpenter@york.ac.uk | |
Organization No | 76 |
Organization acronym | UYRK |
Organization name | University of York |
Organization country/territory | United Kingdom of Great Britain and Northern Ireland |
Address 1 | NCAS, Department of Chemistry |
Address 2 | University of York, Heslington |
Address 3 | York YO10 5DD United Kingdom |
Country/territory | United Kingdom of Great Britain and Northern Ireland |
Tel | |
Fax | |
Last updated date | 2024-04-04 |
Background observation | |
UTC | |
ppm | |
9999-12-31 00:00:00 - 9999-12-31 23:59:59: WMO CO2 X2019 |
|
9999-12-31 00:00:00 - 9999-12-31 23:59:59: Picarro G2401(CRDS) |
|
9999-12-31 00:00:00 - 9999-12-31 23:59:59: 7.5 (m) |
|
4 seconds | |
Calibration of CO2 and CH4 is performed with reference to a single gas standard containing a known amount of CO2/CH4. It is well understood that the instrument behaves linearly within the measurement range. Calibrations are performed once per week by sampling for 30 minutes and the data is applied retrospectively applying a linear drift adjustment. The CO2/CH4 measurements are fully traceable to the international Global Atmospheric Watch (GAW) scale. This means that the gas calibration standards used to calculate the final concentrations are referenced to a common standard held at the NOAA/ESRL laboratory. To ensure this is upheld, the designated world calibration centre for Carbon Monoxide: WCC-Empa (http://www.empa.ch/plugin/template/empa/*/7571), carries out regular audits of the measurements using travelling standards. For more information please see the report from the most recent of these audits: http://www.wmo.int/pages/prog/arep/gaw/documents/CVO_2012.pdf | |
Invalidated data (flag3) includes data when % water content >5%, CO2,CH4,H2O are equal to zero and data when CO2, CH4 <100. Data when CO > 250ppb indicates lab leak and so CH4, CO2 data also invalidated at these times. Flag2 is used as a reduced quality data flag and includes data which is affected by local conditions (wind speed <2 m/s, 350>wind direction >100 degrees) |
|
[Hourly] Hourly: Zero, calibration and invalidated data is removed prior to averaging. Hourly data is obtained by averaging all the data within a 1 hour period for example between 00:00 - 01:00. Standard deviation and number of points used to obtain the average are provided. [Daily] Daily: Zero, calibration and invalidated data is removed prior to averaging. Daily data is obtained by averaging all the data within a 24 hour period for example between 01/01/14 00:00 - 02/01/2014 01:00. Standard deviation and number of points used to obtain the average are provided. [Monthly] Monthly: Zero, calibration and invalidated data is removed prior to averaging. Daily data is obtained by averaging all the data within a 24 hour period for example between 01/01/14 00:00 - 01/02/2014 01:00. Standard deviation and number of points used to obtain the average are provided. |
|
Error Flag = 0 Good data Error Flag = 1 Reduced quality data, also includes local interference data (ws<2/ms, 350>wd >100deg) Error Flag = 2 Below detection limit Error Flag = 3 Invalid or missing data |
|
Valid (background): 0 Valid: 1 Valid: 2 Invalid: 3 |
|
Operational/Reporting | |
Picarro CH4 Measuring range: 0 - 20 ppm Precision (5 sec / 5 min / 60 min, 1σ) <1 ppb /0.5 ppb / 0.3 ppb Max Drift at STP (over 24 hrs / 1 month) 1 ppb /3 ppb Max Uncertainty using Reference Gas (1 hr average, 2σ) < 1 ppb CO2 Measuring range: 0 - 1000 ppm Precision (5 sec / 5 min / 60 min, 1σ) <50 ppb /20 ppb / 10 ppb Max Drift at STP (over 24 hrs / 1 month) 100 ppb /500 ppb Max Uncertainty using Reference Gas (1 hr average, 2σ) < 50 ppb CO2 and CH4 are sampled through 10m of 40mm diameter heated glass inlet. The air then passes through 2m of 1/8" stainless steel tubing and a teflon filter. Initially (2015-Oct 2019) the air was not dried and a the instrument water correction was applied. Since October 2019 the sample has been dried using a nafion drier (PERMAPURE) as we have found that since a return to the manufacturer the water correction is now not acceptable. |
|
Wind direction: 1 Wind speed: 1 Relative humidity: 1 Precipitation amount: 0 Air pressure: 1 Air temperature: 1 Dew point temperature: 0 Sea water temperature: 0 Sea surface water temperature: 0 Sea water salinity: 0 Sea surface water salinity: 0 |
|
Meteorological data may remain as first provided, even when greenhouse gas data are updated. |
Format | Text (WDCGG Data Format Table, WDCGG Meteorological Data Format Table), NetCDF | ||||
---|---|---|---|---|---|
Relation List (Is Part Of) |
All CO2 data contributed to WDCGG by GAW stations and mobiles by 2024-09-24 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2023-09-13 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2022-09-07 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2021-09-08 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2020-09-28 |
||||
Geolocation Point |
|