NO | 27 |
---|---|
Acronym | HKO |
Name | Hong Kong Observatory |
Address 1 | |
Address 2 | Hong Kong Observatory |
Address 3 | 134A Nathan Road, Kowloon, Hong Kong. |
Country/Territory | Hong Kong, China |
Website | http://www.hko.gov.hk |
Name | W.H. Leung |
---|---|
Prefix | Mr. |
whleung@hko.gov.hk | |
Organization No | 27 |
Organization acronym | HKO |
Organization name | Hong Kong Observatory |
Organization country/territory | Hong Kong, China |
Address 1 | |
Address 2 | |
Address 3 | |
Country/territory | Hong Kong, China |
Tel | +852 2926 8421 |
Fax | +852 2375 2645 |
Last updated date | 2024-09-10 |
Name | M.C.Lam |
---|---|
Prefix | Mr. |
mclam@hko.gov.hk | |
Organization No | 27 |
Organization acronym | HKO |
Organization name | Hong Kong Observatory |
Organization country/territory | Hong Kong, China |
Address 1 | |
Address 2 | |
Address 3 | |
Country/territory | Hong Kong, China |
Tel | +852 2926 8024 |
Fax | +852 2375 2645 |
Last updated date | 2024-09-10 |
Background observation | |
UTC+08:00 | |
ppm | |
2010-11-01 00:00:00 - 2021-02-26 23:59:59: WMO CO2 X2007 2021-02-27 00:00:00 - 9999-12-31 23:59:59: WMO CO2 X2019 |
|
2010-11-01 00:00:00 - 2019-12-31 23:59:59: LI-820 from LI-COR Biosciences Ltd.(NDIR) 2020-01-01 00:00:00 - 2021-02-26 23:59:59: LI-840A from LI-COR Biosciences Ltd.(NDIR) 2021-02-27 00:00:00 - 9999-12-31 23:59:59: LI-850 from LI-COR Biosciences Ltd.(NDIR) |
|
9999-12-31 00:00:00 - 9999-12-31 23:59:59: 4 (m) |
|
1 second | |
The LI-820/LI-840A/LI-850 CO2 analyzer is calibrated once every two weeks (based on past experience of the stability of the analyzer; the zero drift as well as the overall system drift is less than 2 ppm in two weeks' time). During the calibration, secondary standard gases of concentration ~400 ppm and ~600 ppm are used for performance check and span calibration of the LI-820/LI-840A/LI-850 respectively. Before the calibration of LI-820/LI-840A/LI-850, the secondary gases are checked for any drift using a LI-7000 CO2 analyzer which is calibrated using the NOAA standard gases. Before the calibration, the LI-820/LI-840A/LI-850 is checked for its performance using the ~400 ppm secondary standard gas and a zero gas. 1. The ~400 ppm standard gas is applied to check for any instrument drift; 2. The zero gas is applied to check for any zero drift; 3. The measured data from steps (1) and (2) are kept for future ref.; 4. Due to the limitation of LI-820/LI-840A/LI-850, only the zero and span calibration are carried out; 5. The LI-820/LI-840A/LI-850 is set zero using the zero gas and then the span point is set using the ~600 ppm secondary standard gas; 6. Finally, the performance of LI-820/LI-840A/LI-850 at around 400 ppm is checked using the ~400 ppm secondary standard gas and the data is kept for future reference. |
|
1. 1-minute average CO2 concentration is calculated from 60 1-second data once every minute; 2. The time series of the raw 1-minute average CO2 concentration is first examined manually to identify any suspicious/erroneous data. Erroneous data are not used for compilation of the hourly and daily mean. 3. Hourly mean CO2 concentration is derived from the 1-minute average; 4. Daily mean CO2 concentration is derived from the hourly mean. |
|
[Hourly] Please refer to the "data processing" section [Daily] Please refer to the "data processing" section [Monthly] Monthly mean are derived from the daily mean |
|
At present, only daily mean CO2 concentration data are submitted and data quality assurance have been carried out at HKO. Hence all data QC flags are 1, meaning data checked OK. | |
Valid: 1 Valid: 2 Invalid: 3 |
|
Operational/Reporting | |
Since November 2010 up to 31 December 2019, instrument used for measuring carbon dioxide concentration was LI-820 from LI-COR Biosciences Ltd. Carbon dioxide concentration in ambient moist air was measured and reported. During the period of 1 January 2020 to 26 February 2021, a newer model LI-840A, was employed to measure carbon dioxide concentration. And since 27 February 2021 onward, LI-850 has been used for the measurement.Both LI-840A and LI-850 sensors measure carbon dioxide as well as water concentration in air, thereby making moisture correction possible through simple post-processing of the data. Carbon dioxide concentration in dry air has been reported from 1 January 2020 onwards. |
|
Wind direction: Wind speed: Relative humidity: Precipitation amount: Air pressure: Air temperature: Dew point temperature: Sea water temperature: Sea surface water temperature: Sea water salinity: Sea surface water salinity: |
|
Meteorological data may remain as first provided, even when greenhouse gas data are updated. |
Format | Text (WDCGG Data Format Table, WDCGG Meteorological Data Format Table), NetCDF | ||||
---|---|---|---|---|---|
Relation List (Is Part Of) |
All CO2 data contributed to WDCGG by GAW stations and mobiles by 2024-09-24 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2023-09-13 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2022-09-07 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2021-09-08 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2020-09-28 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2019-09-19 All CO2 data contributed to WDCGG by GAW stations and mobiles by 2018-10-25 |
||||
Geolocation Point |
|