MMB243N00
-N2O-
| Observation |
| Frequency |
: |
Every 6 hours |
| Method |
: |
GC-ECD |
| Instrument |
: |
Perkin-Elmer |
Perkin-Elmer ƒ°1-ƒ°10 system |
| Description |
: |
The outdoor air was drawn at 2.5 m above the ground surface into a gas sampling loop (2 ml).
|
| A 95% Ar - 5% CH4 mixture was used as the carrier gas. The sample size was 2 ml. Dinitrogen oxide was separated on a 6 ft Porapak QS column at 35 ºC. A reference gas and a sample air were measured in series. |
| Calibration |
| Scale |
: |
The mixing ratio of N2O in the reference air was determined by the method reported in Muramatsu et al.(1982). |
| Description |
: |
The air, which was compressed into 50 l stainless steel container up to 7 kg/cm2 , was used as the reference. |
| Data Processing |
| Unit |
: |
ppb |
| Description |
: |
Only technically false values caused by malfunction of the equipment were removed. |
| Average Procedure |
| Hourly |
: |
|
| Daily |
: |
|
| Monthly |
: |
|
|
|
|