|
|
CPO134S00
-N2O-
| Observation |
| Frequency |
: |
every 15 minutes |
| Method |
: |
GC-ECD |
| Instrument |
: |
SHIMADZU |
|
| Description |
: |
Two freezing units (-4 and -45 ºC respectively) are used in series to dry the ambient air prior to analysis. Sampling is continuous, using a gas chromatograph (GC) with and Electron Capture Detector (ECD). The oven temperature is approximately 40 ºC and the ECD temperature 355 °C. The primary column is a 1/8" x 3 m stainless steel, packed with Porapak N, 80/100 mesh, whilst a pre-column (1/4" x 20 cm s/steel packed with sodium asbestos) is used to extract any CO2. N2 (5.0) is used as carrier gas and the sampling loop (s/steel, 2 mm OD x 2.2 m) has a volume of 7 ml. The data have been recorded at 15 min intervals (96 samples per day) and mixing ratios stored as hourly means (1994 until 2002). From mid-99 onwards, the 15-min intervals have been used to create half-hourly means. These in turn form the basis for the daily and monthly averages.
|
The height of the air intake is 260 m above sea level from the top of a 30 m high mast.
|
| Calibration |
| Scale |
: |
|
| Description |
: |
One working standard is used every hour and multi calibrations involving five (CMDL certified) laboratory standards (range: 286 - 347 ppb N2O) are utilised approximately once every 4 months to verify the working gas concentration. A target gas is also run once daily as a check on instrumental stability.
|
| Data Processing |
| Unit |
: |
ppb |
| Description |
: |
All data with obvious sampling or analytical error are rejected. Furthermore, a statistical filter (moving percentiles) is applied for the selection of background concentrations. |
| Average Procedure |
| Hourly |
: |
|
| Daily |
: |
|
| Monthly |
: |
|
|
|
|